Request more information or a quote
|
|
Hydrolab provides the most accurate Rhodamine WT sensor available on a multiprobe.
Features:
- Ultra-compact size designed specifically for integration into the Hydrolab DS5X, DS5, and MS5
- Available with solid Secondary Standards to provide a quick and simple method to verify the sensor’s stability over time
- The Secondary Standard can be adjusted to correlate to a known dye concentration
- Three auto-selected gain ranges provide measurements from 0.04 to 1000 ppb
Benefits:
- Provides the most accurate measurement of Rhodamine WT because of electronic filtration of ambient light, efficient optical coupling, and quality optical components
- Incredibly fast response time through electronic filtration of ambient light
- Excellent turbidity rejection due to small sample volume design and high quality optical filter
- Cost-optimized for affordability and value
Optical Characteristics: Light Source: Light Emitting Diode Detector: Photodiode Excitation Wavelength: RWT 550 nm Emission Wavelength: RWT 600 nm
Product Data: Minimum Detection Limit: 0.04 ppb Dynamic Range: Low sensitivity: 0.04-1000 ppb Med. sensitivity: 0.04-100 ppb High sensitivity: 0.04-10 ppb Accuracy: +/- 3% for signal level equivalents of 1 ppb rhodamine WT dye or higher using a rhodamine sensor Resolution: 0.01 ppb
Sensor housing: Stainless steel:Standard housing for typical fresh water applications. Titanium option:Corrosion-resistant housing for use in aggressive saline environments such as oceans, bays and estuaries. Consult with your sales representative to determine which sensor best fits your needs. |