What happens if I change altitude after calibrating my DO sensor?

FAQ - Dissolved Oxygen

  1. What is the difference between Hydrolab's Clark Cell technology and competitors?
  2. Do I need a circulator?
  3. Q: What happens if I change altitude after calibrating my DO sensor? A: The Dissolved oxygen readings in mg/L are based on a known saturation (100%) at a known Barometric Pressure; therefore, these readings remain accurate. Percent saturation readings, however, would no longer remain accurate without a new calibration.